Sterilizer and washer-disinfector validation thermocouple
Anville 10B_T_CLASS1_5000_CAP_5A
Part number 10B-T-CLASS1-5000-CAP-5A.
Temperature range -250°C to +250°C.
Accuracy ±0.5°C.
Material Class 1 type T ptfe sheathed.
Length 5m other lengths to order.
Connection Miniature thermocouple plug.
Thermal cycler validation thermocouple for 0.2ml wells
Anville 15A_T0.2_PTFE_1000_5A
Part number 15A-T0.2-PTFE-1000-5A
Temperature range -50°C to +150°C.
Accuracy ±0.5°C.
Material Class 1 type T ptfe sheathed.
Length 1m, other lengths to special order.
Connection Miniature thermocouple plug.
Thermal cycler validation thermocouple for 0.5ml wells
Anville 15A_T0.5_PTFE_1000_5A
Part number 15A-T0.5-PTFE-1000-5A
Temperature range -50°C to +150°C.
Accuracy ±0.5°C.
Material Class 1 type T ptfe sheathed.
Length 1m, other lengths to special order.
Connection Miniature thermocouple plug.
Pressure transducer
For sterilzer and washer-disinfector validation a standard pressure transducer is not adequate as the process temperature is frequently in excess of 80°C. The Ashcroft KXDHT tranducer is designed for continuous operation at process temperature up to 150°C and is much better suited for the purpose. It can be used directly connected to the machine with no cooling loop. Supplied with a fixed cable terminated with a plug suitable for the Series 825 TVI. Other plugs can also be supplied.
Ashcroft KXDHT pressure transducer
Part number KXDHT.
Pressure range -1 bar to 4 bar gauge.
Accuracy ±0.25%.
Span thermal drift ±0.03%/°C.
Repeatability 0.03% full scale
Median temperature range -30°C to 150°C.
Operating temperature range -30°C to 85°C.
Over pressure limit 10 bar.
Signal output 4-20mA.
Power supply 24VDC from Anville data logger.
Process connection G 1/4" male.
Process connection material Stainless steel 3 16Ti.
Connection 3m captive cable terminated with a plug to suit.