01698 535392

Hardware

Temperature Sources

Anville Series 300 (manufactured by Techne Inc.)

For thermocouple sensor calibration of the complete EaziVal system the temperature of the thermocouples must be maintained in a stable state. EaziVal provides multi-point calibration at 'high', 'low' and 'check-point' temperatures with the scale and offset values automatically calculated for each sensor. The Anville 300 bath has an operating range from 40°C to 325°C. The large sensor capacity of three 3mm, three 6mm and one 10mm diameter holes make the unit particularly suitable for thermocouple calibration.

Anville Series 300 Dri-Block® bath


Part number Anville 300

Temperature range 40°C to 325°C

Stability ±0.03°C

Block uniformity ±0.2°C

Accuracy ±0.3°C

Display resolution ±0.1°C

Set point resolution ±0.1°C

Measuring zone 0 to 50mm from base of well

Time to stability 10 minutes (after set temperature is reached

Heating time 9 minutes (from 40°C to 300°C)

Cooling time 18 minutes (from 300°C to 100°C)

Immersion tube 110mm deep

Fixed insert large sensor capacity of three 3mm, three 6mm and one 10mm diameter holes

Dimensions 203.2 x 203.2 x 222.25 mm deep

Power supply 230VAC/110VAC switchable

Weight 4.0kg

Operating ambient 10°C to 40°C

Supplied with two year warranty, fits sterilizer validation kit case or soft case available

Techne Tecal 425H


For thermocouple, PT100 and liquid filled sensor calibration the Tecal 425H provides stable set points with fast heating and cooling. The Tecal 425H controls allow the user to set a number of calibration programmes with the temperature, ramp rate and dwell time stored in the unit.

The Techne Tecal 425H Dri-Block® bath covers a temperature operating range from 40°C to 425°C, has a range of removable inserts. The unit is supplied with a standard insert with two 6mm and two 12mm diameter holes and a soft carrying case.


Techne Tecal 425H Dri-Block® bath


Part number Tecal 425FA

Temperature range 40°C to 425°C

Stability ±0.05°C

Block uniformity ±0.2°C

Accuracy ±0.3°C

Measuring zone 0 to 50mm from base of well

Time to stability 10 minutes (after set temperature is reached)

Heating time 15 minutes (from 40°C to 400°C)

Cooling time 25 minutes (from 400°C to 100°C

Immersion tube 114mm deep x 38mm diameter

Standard insert 2x6mm, 2x12mm diameter holes

Power supply 230VAC

Weight 9kg

Operating ambient 10°C to 40°C

Supplied with two year warranty and soft carry case

Techne Tecal 140F and 425F


The Techne 140F and 425F dry-block baths provide stability with fast heating and cooling enabling rapid setting of the three temperature set points.

The Tecal 140F dry-block bath covers a temperature operating range from -20°C to 140°C and has a range of removable inserts. The Tecal 425F dry-block bath covers a temperature operating range from -40°C to 425°C and has a range of removable inserts. The units are supplied with a standard insert with five 6mm diameter holes and a soft carry case.


Techne Tecal 140F Dri-Block® bath


Part number Tecal 140F

Temperature range -20°C to 140°C

Stability ±0.05°C

Block uniformity ±0.2°C

Accuracy ±0.3°C

Display resolution ±0.01°C

Set point resolution ±0.1°C

Measuring zone 0 to 50mm from base of well

Time to stability 10 minutes (after set temperature is reached)

Heating time 5 minutes (from 20°C to 100°C)

Cooling time 9 minutes (from 100°C to 0°C)

Immersion tube 114mm deep x 38mm diameter

Standard insert 5x6mm diameter holes

Dimensions 207mm wide x 273mm high x 289mm deep

Power supply 230VAC

Weight 11kg

Operating ambient 10°C to 40°C

Supplied with two year warranty and soft carry case

Techne Tecal 425F Dri-Block® bath


Part number Tecal 425FA

Temperature range 40°C to 425°C

Stability ±0.03°C

Block uniformity ±0.2°C

Accuracy ±0.3°C

Display resolution ±0.01°C

Set point resolution ±0.1°C

Measuring zone 0 to 50mm from base of well

Time to stability 10 minutes (after set temperature is reached)

Heating time 12 minutes (from 40°C to 400°C)

Cooling time 21 minutes (from 400°C to 100°C)

Immersion tube 114mm deep x 38mm diameter

Standard insert 5x6mm diameter holes

Dimensions 170mm wide x 270mm high x 255mm deep

Power supply 230VAC

Weight 6.3kg

Operating ambient 10°C to 40°C

Supplied with two year warranty and soft carry case

Techne thermoregulator and liquid bath


Calibration of the PL-105 and SterilProbe data loggers requires total immersion of the data logger in a liquid bath maintained at a stable, known temperature. A techne 8 litre liquid bath fitted with a Techne TE-10D thermo-regulator has a calibration temperature range from 40°C to 120°C. The liquid can be either water if temperatures do not exceed 95°C (for washer-disinfector calibration) or silicone oil for temperatures up to 120°C.

The bath temperature can be measured with a UKAS calibrated Anville series 275 reference thermometer.


Techne thermoregulator and liquid bath


Part number FBATH08 Stainless steel bath with 8 litre capacity.

Part number FFLAT08 Stainless steel lid.

Part number FTE10DDC Thermo-regulator with digital display of set point, operating range 40°C to 120°C. The bridge mounting plate has an entry port of 6.5mm diameter for the 275 Pt100 sensor.