01698 535392

Hardware

Thermocouples and Transducers

For sterilizer, washer-disinfector and thermal cycler validation Anville offer a range of thermocouples for use with the Series 825 TVI. Constructed using the best quality material for accurate and repeatable measurements the thermocouples are manufactured using class type T material in ptfe sheathing.

Sterilizer and washer-disinfector validation thermocouple


Anville 10B_T_CLASS1_5000_CAP_5A

Part number 10B-T-CLASS1-5000-CAP-5A.

Temperature range -250°C to +250°C.

Accuracy ±0.5°C.

Material Class 1 type T ptfe sheathed.

Length 5m other lengths to order.

Connection Miniature thermocouple plug.

Thermal cycler validation thermocouple for 0.2ml wells


Anville 15A_T0.2_PTFE_1000_5A

Part number 15A-T0.2-PTFE-1000-5A

Temperature range -50°C to +150°C.

Accuracy ±0.5°C.

Material Class 1 type T ptfe sheathed.

Length 1m, other lengths to special order.

Connection Miniature thermocouple plug.

Thermal cycler validation thermocouple for 0.5ml wells


Anville 15A_T0.5_PTFE_1000_5A

Part number 15A-T0.5-PTFE-1000-5A

Temperature range -50°C to +150°C.

Accuracy ±0.5°C.

Material Class 1 type T ptfe sheathed.

Length 1m, other lengths to special order.

Connection Miniature thermocouple plug.

Pressure transducer


For sterilzer and washer-disinfector validation a standard pressure transducer is not adequate as the process temperature is frequently in excess of 80°C. The Ashcroft KXDHT tranducer is designed for continuous operation at process temperature up to 150°C and is much better suited for the purpose. It can be used directly connected to the machine with no cooling loop. Supplied with a fixed cable terminated with a plug suitable for the Series 825 TVI. Other plugs can also be supplied.


Ashcroft KXDHT pressure transducer


Part number KXDHT.

Pressure range -1 bar to 4 bar gauge.

Accuracy ±0.25%.

Span thermal drift ±0.03%/°C.

Repeatability 0.03% full scale

Median temperature range -30°C to 150°C.

Operating temperature range -30°C to 85°C.

Over pressure limit 10 bar.

Signal output 4-20mA.

Power supply 24VDC from Anville data logger.

Process connection G 1/4" male.

Process connection material Stainless steel 3 16Ti.

Connection 3m captive cable terminated with a plug to suit.