01698 535392

Thermal Validation Hardware

Series 300 Temperature Calibrator

For thermocouple sensor calibration of the complete EaziVal system the temperature of the thermocouples must be maintained in a stable state. EaziVal provides multi-point calibration at 'high', 'low' and 'check-point' temperatures with the scale and offset values automatically calculated for each sensor. The Anville 300 bath has an operating range from 40°C to 325°C. The large sensor capacity of three 3mm, three 6mm and one 10mm diameter holes make the unit particularly suitable for thermocouple calibration.

Anville Series 300 bath


Part number Anville 300

Temperature range 40°C to 325°C

Stability ±0.03°C

Block uniformity ±0.2°C

Accuracy ±0.3°C

Display resolution ±0.1°C

Set point resolution ±0.1°C

Measuring zone 0 to 50mm from base of well

Time to stability 10 minutes (after set temperature is reached

Heating time 9 minutes (from 40°C to 300°C)

Cooling time 18 minutes (from 300°C to 100°C)

Immersion tube 110mm deep

Fixed insert large sensor capacity of three 3mm, three 6mm and one 10mm diameter holes

Dimensions 203.2 x 203.2 x 222.25 mm deep

Power supply 230VAC/110VAC switchable

Weight 4.0kg

Operating ambient 10°C to 40°C

Supplied with two year warranty, fits sterilizer validation kit case or soft case available