01698 535392

Prodigy

Sterilizer and Washer-Disinfector Monitoring

Sterilizer Monitoring and Washer Disinfector Monitoring Systems

For sterilizer and washer-disinfector independent cycle monitoring an accurate, reliable measurement system is required. The Anville systems provide truly independent cycle monitoring and can be used on both single cabinet and tunnel washer-disinfectors and sterilizers.

The system can be used immediately - no programming is required. More flexible than a chart recorder, the Series 800 with Prodigy software sets a new standard with its ease of system operation and configuration, providing data acquisition, data display, alarm monitoring and end-of-cycle report generation

Overview Display


This display is a schematic of the department. After the software has started the overview display will be shown. This display is used to navigate to the displays for the individual machines being monitored and will be also selected by clicking an ‘Overview’ button from any other user-selected display.


Sterilizer & Washer-disinfector Display


Each sterilizer and washer-disinfector has a separate display page with a trend graph; machine status, cycle number and current temperature, pressure, flow and conductivity data etc.


Report Display & Print-out


A "proof of process" report can be printed out automatically at the end of each cycle. A list of all available cycles for that autoclave sterilizer or washer-disinfector can be viewed and printed out on manual command.


Data Archiving


The data archiving facility allows long term secure archiving of each batch cycle from each sterilizer and washer-disinfector.


Hardware


Anville Instruments sterilizer and washer disinfector monitoring systems use the same hardware base as the Anville Series 825 thermal validation interface.


Anville Series 800


Each Series 800 monitors temperature, pressure, conductivity, detergent flow, machine status sensors from a single sterilizer or washer-disinfector.
The Series 800/WD has eight analogue inputs, two status inputs and controls two relay outputs. The analogue inputs can be configured to monitor PT100 temperature sensors, pressure transducers, conductivity transmitters etc.

The Series 800/ST has four dedicated type T thermocouple inputs and four other analogue inputs, which can be configured to monitor pressure transducers conductivity transmitters etc., two status inputs and controls two relay outputs.

Supplied in a waterproof housing, the Series 800 can be mounted adjacent to the sterilizer or washer-disinfector, reducing sensor lead lengths and installation problems. Connection to the host computer is by standard serial communication, or Ethernet.